Antireflux valves in patient‐controlled analgesia

M. T. Kluger, H. Owen

    Research output: Contribution to journalArticlepeer-review

    13 Citations (Scopus)

    Abstract

    Antireflux valves are widely used in conjunction with patient‐controlled analgesia devices. It is important to appreciate the limitations and dangers of these systems. They can achieve a potential ‘stored volume’ if occluded and they may, as part of the administration set, retard fluid administration. Seven antireflux systems currently available were tested in conjunction with three patient‐controlled analgesia pumps. The systems’ volume, time to occlusion alarm and flow rates were measured. Results showed that the sets with low stored volumes were less efficient as administration sets. A potentially dangerous bolus could result after release of occlusion if sets with large stored volumes were used in conjunction with pumps that utilised concentrated solutions of opioid. This study has identified the ideal antireflux valve system.

    Original languageEnglish
    Pages (from-to)1057-1061
    Number of pages5
    JournalAnaesthesia
    Volume45
    Issue number12
    DOIs
    Publication statusPublished - Dec 1990

    Keywords

    • Equipment; infusion systems, patient‐controlled

    Fingerprint

    Dive into the research topics of 'Antireflux valves in patient‐controlled analgesia'. Together they form a unique fingerprint.

    Cite this