Automated bone screw tightening to adaptive levels of stripping torque

Karen J. Reynolds, Aaron A. Mohtar, Tammy M. Cleek, Melissa K. Ryan, Trevor C. Hearn

    Research output: Contribution to journalArticle

    Abstract

    Objective: To use relationships between tightening parameters, related to bone quality, to develop an automated system that determines and controls the level of screw tightening. Methods: An algorithm relating current at head contact (I HC) to current at construct failure (I max) was developed. The algorithm was used to trigger cessation of screw insertion at a predefined tightening level, in real time, between head contact and maximum current. The ability of the device to stop at the predefined level was assessed. Results: The mean (±SD) current at which screw insertion ceased was calculated to be [51.47 ± 9.75% × (I max - I HC)] + I HC, with no premature bone failures. Conclusions: A smart screwdriver was developed that uses the current from the motor driving the screw to predict the current at which the screw will strip the bone threads. The device was implemented and was able to achieve motor shut-off and cease tightening at a predefined threshold, with no premature bone failures.

    Original languageEnglish
    Pages (from-to)321-325
    Number of pages5
    JournalJournal of Orthopaedic Trauma
    Volume31
    Issue number6
    DOIs
    Publication statusPublished - Jun 2017

    Keywords

    • adaptive screw tightening
    • cancellous bone screw
    • computer-assisted surgery

    Fingerprint Dive into the research topics of 'Automated bone screw tightening to adaptive levels of stripping torque'. Together they form a unique fingerprint.