Impedance nanopore biosensor: influence of pore dimensions on biosensing performance

Krishna Kant, Jingxian Yu, Craig Priest, Joseph Shapter, Dusan Losic

    Research output: Contribution to journalArticlepeer-review

    38 Citations (Scopus)

    Abstract

    Knowledge about electrochemical and electrical properties of nanopore structures and the influence of pore dimensions on these properties is important for the development of nanopore biosensing devices. The aim of this study was to explore the influence of nanopore dimensions (diameter and length) on biosensing performance using non-faradic electrochemical impedance spectroscopy (EIS). Nanoporous alumina membranes (NPAMs) prepared by self-ordered electrochemical anodization of aluminium were used as model nanopore sensing platforms. NPAMs with different pore diameters (25-65 nm) and lengths (4-18 μm) were prepared and the internal pore surface chemistry was modified by covalently attaching streptavidin and biotin. The performance of this antibody nanopore biosensing platform was evaluated using various concentrations of biotin as a model analyte. EIS measurements of pore resistivity and conductivity were carried out for pores with different diameters and lengths. The results showed that smaller pore dimensions of 25 nm and pore lengths up to 10 μm provide better biosensing performance.

    Original languageEnglish
    Pages (from-to)1134-1140
    Number of pages7
    JournalAnalyst
    Volume139
    Issue number5
    DOIs
    Publication statusPublished - 3 Feb 2014

    Fingerprint

    Dive into the research topics of 'Impedance nanopore biosensor: influence of pore dimensions on biosensing performance'. Together they form a unique fingerprint.