Respiratory gating of Anatomical optical coherence tomography images of the human airway

Robert A. Mclaughlin, Julian J. Armstrong, Sven Becker, Jennifer H. Walsh, Arpit Jain, David R. Hillman, Peter R. Eastwood, David D. Sampson

Research output: Contribution to journalArticlepeer-review

21 Citations (Scopus)

Abstract

Anatomical optical coherence tomography (aOCT) is a longrange endoscopic imaging modality capable of quantifying size and shape of the human airway. A challenge to its in vivo application is motion artifact due to respiratory-related movement of the airway walls. This paper represents the first demonstration of respiratory gating of aOCT airway data, and introduces a novel error measure to guide appropriate parameter selection. Results indicate that at least four gates per respiratory cycle should be used, with only minor improvements as the number of gates is further increased. It is shown that respiratory gating can substantially improve the quality of aOCT images and reveal events and features that are otherwise obscured by blurring.

Original languageEnglish
Pages (from-to)6568-6577
Number of pages10
JournalOptics Express
Volume17
Issue number8
DOIs
Publication statusPublished - 13 Apr 2009
Externally publishedYes

Fingerprint Dive into the research topics of 'Respiratory gating of Anatomical optical coherence tomography images of the human airway'. Together they form a unique fingerprint.