Abstract
Accurate and rapid diagnostic tests are a critical component for the early diagnosis of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) and of the overall control strategy for the current pandemic. Nucleic acid amplification tests are the gold standard for diagnosis of acute SARS-CoV-2 infection, and many real-time PCR diagnostic assays have been developed. Mutations that occur within the primer/probe binding regions of the SARS-CoV-2 genome can negatively impact the performance of diagnostic assays. Here, we report two single-point mutations in the N gene of SARS-CoV-2 associated with N gene target detection failures in the Cepheid Xpert Xpress SARS-CoV-2 assay, the first a C to T mutation at position 29197, found in five patients, and the second a C to T mutation at position 29200, found in eight patients. By sequencing the Xpert amplicons, we showed both mutations to be located within the amplified region of the Xpert N gene target. This report highlights the necessity for multiple genetic targets and the continual monitoring and evaluation of diagnostic assay performance.
| Original language | English |
|---|---|
| Article number | e01494-21 |
| Number of pages | 7 |
| Journal | Microbiology Spectrum |
| Volume | 9 |
| Issue number | 3 |
| DOIs | |
| Publication status | Published - Dec 2021 |
| Externally published | Yes |
Keywords
- Cepheid Xpert
- Diagnostic assay
- Mutation
- N gene
- SARS-CoV-2