Abstract
BACKGROUND: Patients with acute coronary syndromes (ACS) have been shown to have a local increase in culprit lesion temperature of at least 0.5 degrees Celcius using a specialized thermography catheter. However, this device is unique, not clinically available and unable to provide information other than temperature. The 0.014-inch Radi™ PressureWire® XT (RPW) contains a high-sensitivity thermistor in the sensor that has a sensitivity of 0.1 degrees Celcius. We evaluated the ability of the RPW to detect an increase in plaque temperature in patients with ACS. METHODS AND RESULTS: Patients with ACS and a culprit lesion of > 70% stenosis and who were candidates for percutaneous coronary intervention (PCI) were eligible (n = 20). Fractional flow reserve (FFR) post-adenosine administration and temperature estimations were performed prior to PCI. All demographic data are presented as mean ± SD, and temperature data (using delta temperature from baseline) as mean ± SEM. Fifteen men and 5 women were enrolled (age 59.5 ± 11.6 years). The FFR pre-PCI was 0.65 ± 0.06, consistent with hemodynamically significant stenoses. The baseline delta temperature was 0.00 ± 0.01 degrees Celcius. The delta temperature at the culprit lesion was -0.02 ± 0.01 degrees Celcius (p = NS), a result below the resolution of the thermistor. Post-PCI, the FFR was 0.95 ± 0.01 (p < 0.0001). CONCLUSIONS: The RPW was unable to detect any temperature increase in patients with ACS.
Original language | English |
---|---|
Pages (from-to) | 528-531 |
Number of pages | 4 |
Journal | Journal of Invasive Cardiology |
Volume | 18 |
Issue number | 11 |
Publication status | Published - Nov 2006 |